Legacy software or instruments may experience compatibility issues. Please receive support from customer service in such cases.
Please note:
Some optional features require separate hardware parts. Please make sure to assemble them before software installation!
go onThe installation software is deployed on a flash drive. It‘s stored
in the envelope that contains the manual and certificate.
Express Setup
Automatic installation, simply follow the prompts.
Custom Setup
For advanced users desiring custom settings. Follow the prompts.
Some Laser Spectrum Analyzers have an external power supply. Please connect the power supply before connecting the USB cable (included in shipment). Please avoid using USB cables longer than 3 meters – that might cause communication protocol difficulties with Windows.
Please note: A few nJ (10 nW @ 1 s exposure) or even less are usually enough to achieve a measurement. Take special care when working with pulsed lasers to avoid exceeding damage threshold of optical fibers.
Suitable patchcords:
All laser spectrum analyzers feature FC/PC sockets.
For singlemode fibers: Use the FC/APC connector at the
laser and the FC/PC connector at the laser spectrum analyzer.
Your measurement results will be impaired when you apply
FC/APC connectors to a laser spectrum analyzer input.
Simply run the desktop shortcut.
Depending on your light source adjust the software settings in the graphical user interface.
A) Select preferred unit
B) Select pulsed or continuous
C) Select Adjust exposure manually or select “Automatic”
D) Start the measurement
E) Click on “Analysis” to display the laser spectrum
instead of the raw diffraction patterns.
Click on “Linewidth” to display the FWHM.
Don't hesitate to contact our service team for support.